Floating reference position-based correction method for near-infrared spectroscopy in long-term glucose concentration monitoring
点击次数:
- 所属单位:Tianjin University
- 教研室:Bio optics and Applied Spectroscopy Laboratory
- 发表刊物:Journal of Biomedical Optics
- 刊物所在地:America
- 项目来源:New Hemanalysis Instrument-Noninvasive Blood Glucose Detector(新型血液分析仪器—无创血糖检测仪)
- 关键字:near-infrared spectroscopy;drift correction;scattering medium;blood glucose
- 摘要:We present a floating reference position (FRP)-based drift correction method for near-infrared (NIR) spectroscopy-based long-term blood glucose concentration (BGC) monitoring. Previously, we reported that it is difficult to quantify the systematic drift caused by the fluctuation of incident light intensity at different source–detector (SD) separations based on the absolute FRP change. We use the relative FRP change as a baseline reference to quantitatively characterize the signal drift at different SD separations.