Abstract:Seventeen cases of glucose tolerance test (GTT) on healthy and diabetic volunteers were conducted to validate the strategy. The human experiment results clearly show that the expected blood glucose changes have been detected at 1550 nm. The average correlation coefficient of the 17 cases of GTT between light signal and glucose reference reaches 0.84. The proposed measurement strategy is verified feasible for the glucose detecting in vivo. The strategy provides references to further studies and product developments for the NIR spectroscopy-based glucose measurement.
Note:DOI: 10.1177/00037028221092474
First Author:Liu Jin; Han Tongshuai
Indexed by:Applied Research
Correspondence Author:Xu Kexin
Volume:76
Issue:9
Translation or Not:no